![]() Troubleshooting, or performing other maintenance on the product or its controls.ģ General Recommendations The refrigeration system is designed to maintain cryogenic temperatures with safety only when the freezer is used for storage. The sublids are necessary to maintain correct temperature, moisture control, and economy of operation. CAUTION! Do not discard the sublids from chest-style units. When submitting a claim for shipping damage, request that the carrier inspect the shipping container and equipment. Do not return goods without written authorization. If you find any damage, keep the packing materials and immediately report the damage to the carrier. If there is no exterior damage, unpack and inspect the equipment within five days of delivery. If exterior damage is present, carefully unpack and inspect the unit and all accessories for damage. CAUTION! IFailure to follow these procedures or overloading the unit may cause undue stress on the compressors or jeopardize user product safety.Ĥ Pre-Installation 4.1 Unpacking At delivery, examine the exterior for physical damage while the carrier’s representative is present. Allow the freezer to operate at the desired temperature for a minimum of 12 hours before loading. The setpoint should be no warmer than -50✬.